working principle
Adopting micro melting technology and introducing aviation application technology, using high-temperature glass to melt micro machined silicon varistor strain gauges
On the stainless steel isolation membrane. The glass bonding process avoids the effects of temperature, humidity, mechanical fatigue, and medium on the adhesive and
The influence of materials has improved the long-term stability performance of sensors in industrial environments, while also avoiding sensing
The P-N junction effect phenomenon that occurs in traditional micro machining manufacturing process engineering.
Technical Specifications
·Measurement range
Relative pressure: minimum measurement range(0~0.1)MPa,Maximum measurement range(0~70)MPa
Negative relative pressure: minimum measurement range (-0.1~0.1)MPa; Maximum measurement range (-0.1~70)MPa
·Measurement accuracy
1%FS (including linear, hysteresis, and repeatability)
·Long term stability
Every year ≤±0.25%FS
·Allow ambient temperature
-40℃~+85℃
·Allowable medium temperature
-40℃~+125℃
·Allow storage temperature
-40℃~+125℃
·Temperature influence
1.0%, 0.5% accuracy: compensating temperature from 0 ℃ to+50 ℃, with a full temperature range impact of ≤± 0.75%
0.25% accuracy: compensation temperature -20 ℃~+60 ℃, full temperature range impact ≤± 0.5%
Or special requirements agreed upon
·Signal output
(4-20) mA analog type (two-wire system)
(0.5-4.5) V DC (three wire system)
(0-10) V DC (three wire system)
·Process Connection
External thread G1/4
·Anti vibration
20g (20~200Hz)
·Impact resistance
200g/11ms
·Working power supply
(4-20) mA output type (12-30) V DC;(0.5-4.5) V DC output type:5V DC
·Material of the part in contact with the medium
17-4PH stainless steel
·housing material
Stainless Steel 304
·Protection level
Aviation plug IP65; Overall cable IP65;
·Electromagnetic compatibility interference
EN61000-6-1,EN61000-6-310V/m
·Electrical connection
Overall cable type; Aviation plug-in type
·Pressure cycle
>107 full-scale cycles
·Other parameters
Overload pressure 2 times the rated pressure
Breaking pressure 5 times rated pressure
Explosion proof Ex d IIC T6 Gb
Load impedance: Current output RLMAX [Ω] ≤ (US-12V)/0.02A
(RLMAX [Ω] ≤ 600 Ω at 24V, RLMAX [Ω] ≤ 900 Ω at 30V)
Voltage output RLmin [Ω] ≥ 5 k Ω
Exterior structure
When installing instruments, it is forbidden to rotate the aviation plug with hands or tools!
electrical connection
In environments with strong interference, in order to ensure the normal operation of instruments, they should be reliably grounded according to relevant instrument grounding technical specifications!